Radiotherapy and Oncology. 1996 Jan;38(1) :61-68
PMID: ISI:A1996TX03200009   
Back to previous list
Abstract
A stereotactic radiosurgery and radiotherapy (SRS/SRT) system on a dedicated Varian Clinac-600SR linear accelerator with Brown-Reberts-Wells and Gill-Thomas-Cosman relocatable Frames along with the Radionics (RSA) planning system is evaluated. The Clinac-600SR has a single 6-MV beam with the same beam characteristics as that of the mother unit, the Clinac-600C. The primary collimator is a fixed cone projecting to a 10-cm diameter at isocenter. The secondary collimator is a heavily shielded cylindrical collimator attached to the face plate of the primary collimator. The tertiary collimation consists of the actual treatment cones. The cone sizes vary from 12.5 to 40.0 mm diameter. The mechanical stability of the entire system was verified. The variations in isocenter position with table, gantry, and collimator rotation were found to be <0.5 mm with a compounded accuracy of less than or equal to 1.0 mm. The radiation leakage under the cones was < 1(a)` o measured at a depth of 5 cm in a phantom. The beam profiles of all cones in the x and y directions were within +/- 0.5 mm and match with the physical size of the cone. The dosimetric data such as tissue maximum ratio, off-axis ratio, and cone factor were taken using film, diamond detector, and ion chambers. The mechanical and dosimetric characteristics including dose linearity of this unit are presented and found to be suitable for SRS/SRT. The difficult in absolute dose measurement For small cone is discussed.
Notes
Times Cited: 9 English Article TX032 RADIOTHER ONCOL