Stathakis S , Li JS , Ma CCM
Monte Carlo determination of radiation-induced cancer risks for prostate patients undergoing intensity-modulated radiation therapy
Journal of Applied Clinical Medical Physics. 2007 ;8(4) :14-27
PMID: ISI:000253440600003   
Back to previous list
Abstract
The application of intensity-modulated radiation therapy (IMRT) has enabled the delivery of high doses to the target volume while sparing the surrounding normal tissues. The drawbacks of intensity modulation, as implemented using a computer-controlled multileaf collimator (MLC), are the larger number of monitor units (MUs) and longer beam-on time as compared with conventional radiotherapy. Additionally, IMRT uses more beam directions-typically 5-9 for prostate treatment-to achieve highly conformal dose and normal-tissue sparing. In the present work, we study radiation-induced cancer risks attributable to IMRT delivery using MLC for prostate patients. Whole-body computed tomography scans were used in our study to calculate (according to report no. 116 from the National Council on Radiation Protection and Measurements) the effective dose equivalent received by individual organs. We used EGS4 and MCSIM to compute the dose for IMRT and three-dimensional conformal radiotherapy. The effects of collimator rotation, distance from the treatment field, and scatter and leakage contribution to the whole-body dose were investigated. We calculated the whole-body dose equivalent to estimate the increase in the risk of secondary malignancies. Our results showed an overall doubling in the risk of secondary malignancies from the application of IMRT as compared with conventional radiotherapy. This increase in the risk of secondary malignancies is not necessarily related to a relative increase in MUs. The whole-body dose equivalent was also affected by collimator rotation, field size, and the energy of the photon beam. Smaller field sizes of low-energy photon beams (that is, 6 MV) with the MLC axis along the lateral axis of the patient resulted in the lowest whole-body dose. Our results can be used to evaluate the risk of secondary malignancies for prostate IMRT patients.
Notes
ISI Document Delivery No.: 266MR Times Cited: 0 Cited Reference Count: 32 Cited References: *ICRP, 1991, 1990 REC INT COMM RA *NCRP, 1993, 116 NCRP BEY P, 2000, INT J RADIAT ONCOL, V48, P513 BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165 BLETTNER M, 1991, STAT MED, V10, P1511 BOICE JD, 1985, J NATL CANCER I, V74, P955 BOICE JD, 1987, J NATL CANCER I, V79, P1295 BOYER AL, 2001, INT J RADIAT ONCOL, V51, P880 BRENNER DJ, 2000, CANCER, V88, P398 CURTIS RE, 1994, J NATL CANCER I, V86, P1315 DAY N, 1983, IARC SCI PUBLICATION, V52 DENG J, 2000, PHYS MED BIOL, V45, P411 DORR W, 2002, J RADIOL PROT, V22, A117 FOLLOWILL D, 1997, INT J RADIAT ONCOL, V38, P667 HALL EJ, 2003, INT J RADIAT ONCOL, V56, P83 HANKS GE, 1998, INT J RADIAT ONCOL, V41, P501 HANKS GE, 2002, INT J RADIAT ONCOL, V54, P427 KAPUR A, 1997, P 12 INT C US COMP R, P157 KRY SF, 2005, INT J RADIAT ONCOL, V62, P1204 LYONS JA, 2000, UROLOGY, V55, P85 MA CM, 1997, MED PHYS, V24, P401 MA CM, 2004, P 14 INT C US COMP R, P515 MOVSAS B, 1998, INT J RADIAT ONCOL, V41, P251 NELSON WR, 1985, 265 SLAC POLLACK A, 2000, INT J RADIAT ONCOL, V48, P507 POLLACK A, 2002, INT J RADIAT ONCOL, V53, P1097 PRICE RA, 2003, J APPL CLIN MED PHYS, V4, P231 ROGERS DWO, 1995, MED PHYS, V22, P503 UPTON AC, 1961, CANCER RES, V21, P717 WEBB S, 2000, PHYS MED BIOL, V45, P1715 YANG J, 2004, PHYS MED BIOL, V49, P2657 ZELEFSKY MJ, 1998, INT J RADIAT ONCOL, V41, P491 Stathakis, Sotirios Li, Jinsheng Ma, Charlie C. M. MULTIMED INC; 66 MARTIN ST, TORONTO, ON L9T 2R2, CANADA