Zhu TC , Stathakis S , Clark JR , Feng W , Georg D , Holmes SM , Kry SF , Ma CC , Miften M , Mihailidis D , Moran JM , Papanikolaou N , Poppe B , Xiao Y
Report of AAPM Task Group 219 on independent calculation-based dose/MU verification for IMRT
Med Phys. 2021 Jul 2
Back to previous list
Abstract
Independent verification of the dose per monitor unit (MU) to deliver the prescribed dose to a patient has been a mainstay of radiation oncology quality assurance (QA). We discuss the role of secondary dose/MU calculation programs as part of a comprehensive QA program. This report provides guidelines on calculation-based dose/MU verification for intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT) provided by various modalities. We provide a review of various algorithms for "independent/second check" of monitor unit calculations for IMRT/VMAT. The report makes recommendations on the clinical implementation of secondary dose/MU calculation programs; on commissioning and acceptance of various commercially available secondary dose/MU calculation programs; on benchmark QA and periodic QA; and on clinically reasonable action levels for agreement of secondary dose/MU calculation programs.
Notes
2473-4209 Zhu, Timothy C Stathakis, Sotiris Clark, Jennifer R Feng, Wenzheng Georg, Dietmar Holmes, Shannon M Kry, Stephen F Ma, Chang-Ming Charlie Miften, Moyed Mihailidis, Dimitris Moran, Jean M Papanikolaou, Niko Poppe, Bjorn Xiao, Ying Journal Article United States Med Phys. 2021 Jul 2. doi: 10.1002/mp.15069.